01 03 0402
●Power source: DC 3V( 2 Section 7 alkaline
batteries)
●Measurement range: 32.0℃ ~43.0℃
●Display resolution: 0.1℃
●Measurement accuracy: ±0.2℃ in the range of
35.0℃~42.0℃
±0.3℃in the range of 32.0℃~34.9℃ and 42.1℃~
43.0℃
●Electric shock protection: the device is supplied by
internal power
●Applied part: BF Type
●Running mode: continuous running
●Temperature units: ℃/℉
●Degrees of protection against ingress liquid: IP20
(solid foreign matter whose diameter are greater than
or equal to 12.5mm can be prevented from entering,
no protection against ingress liquid )
●Safety classification: the device that cannot be used
in the presence of flammable anesthetic gases mixed
with air or oxygen or nitrous oxide
●Product size: 149×38×42(mm)
●Product weight:about 68g (not including cell)
This product complies with the Electromagnetic
Compatibility (EMC) standard for safe operation of
electrical equipment for medical use and
YY0505-2012.
The standard of YY0505-2012 (item 6.8.2.201)
stipulates that users should be provided with detailed
information related to EMC environment for safe
operation of equipment. And the following is the
description of technical instructions related to EMC.
(For details, please refer to YY0505-2012.)
The Electromagnetic Compatibility (EMC) standard
represented by YY0505 set a criterion for the safe
operation of electrical equipment for medical use.
The standard stipulates that the interference of the
noise generated by the equipment to other devices
should be controlled within a certain range, so is the
electromagnetic wave dry disturbance emitted by
other devices (mobile phones, etc.) .
1. Definition of EMC (electromagnetic compatibility)
EMC refers to the ability to satisfy the following two
requirements.
●No unallowable electromagnetic interference noise
will be emitted to other nearby electronic devices.
(Radiation)
●The device can work normally in the
electromagnetic environment with noise and other
interference from other electronic devices. (Noise
immunity)
2. Technical specification on EMC (electromagnetic
compatibility)
Electrical equipment for medical use requires specific
instructions for EMC and should be operated in
accordance with the following EMC information .
This device requires special instructions on
electromagnetic compatibility (EMC). Please use the
device in accordance with the EMC information
described in this manual.
Portable and wireless radio frequency (RF)
communication devices may affect the device.
●When using this device, do not place it adjacent to
or stack it with other devices.
11.The Use of Electromagnetic Environment
Guidance
1. The measurement result is only for reference,
which is not a substitute for a physician's diagnosis. It
is very dangerous to self-judge and treat only based
on the measurement result. Please follow the
doctor's instructions.
2. Please put the battery out of the reach of children,
otherwise it is dangerous.
3. When the product is not used for a long time (more
than 3 months), remove the battery from device to
prevent the battery leakage.
4. It is forbidden to immerse the infrared thermometer
in any liquid, and It is forbidden to use it for a long time
under too high or low temperature condition. No
collisions, drops and mixing with sharp objects.
5. Do not put the battery close to the fire or into the fire
to avoid the battery explosion. Do not use the battery
when it leaks or molds; when discarding batteries or
this product, it should be following local regulations to
avoid contamination.
6. This product contains sensitive electronic
components, should avoid strong electromagnetic
interference.
7. Please read this instruction carefully before use to
confirm that the battery is installed.
8. Do not disassemble, repair or remould this device,
which may lead to measurement error or machine
malfunction.
9. This product is not waterproof and should be used
carefully to prevent liquid from entering the product.
When the product gets wet due to contact with steam,
use after it become dry or gently wipe with a soft dry
cloth or cotton ball, otherwise it will cause
measurement errors.
●measurement environment:
8. Measurement and transport storage
environment
●Installing the battery
The device is supplied with 2
Section 7 alkaline batteries. Push
the battery cover downward and
load the battery into the battery
compartment. At this time, the
device will start self-inspection.
Pay attention to the positive and
negative poles instruction in the
5. Installation and usage
AAA
AAA
●This device is used to measure the body
temperature by passively sensing infrared heat
radiation collected from the human body's
forehead, which can be used for families and
medical departments to measure the body
temperature .
●Intended use and application scope: This device
is used to measure the forehead temperature of the
human body, which is applicable to infants, children
and adults.
●Contraindication: None.
2. Device structure and composition
1. Device use and applicant scope
Symbols Implication
Type BF application part
Electronic information product pollution
control mark environmental protection
use period is 10 years. Consumables
are not included
Caution! Consult instruction of product
Paper recycling
This side up
Fragile
Non-rainproof
●LCD area
●Appendix:
Instructions, Quick Operation Guide, 2 Section 7
Alkaline Batteries.
6. Product maintance and upkeep method
As the product is reusable device, please pay
attention to cleaning and disinfection after use.
If the product is dirty, please keep the sensor and
probe cavity clean, otherwise the measurement
accuracy will be affected.
How to clean sensor and probe cavity:gently wipe the
inner cavity or sensor mirror with a clean soft cloth or
Phenomenon
of breakdown
The measurement
temperature is lower
than 32℃, which is
beyond the
measurement range
Please re-measure
following the
product instruction
The measurement
temperature is
higher than 43℃,
which is beyond the
measurement range
Change the cell
No
display or
abnormal
display
The positive and
negative poles of the
cell are reversed or the
device is abnormal
7. Common failures and troubleshooting
methods
Please measure
under the 16℃~35℃
environment
temperature
The
screen
display
" "
The
screen
display
" "
The screen
display
" "
The power of cell is
shortage
Re-install the cell, or
contact the dealer
The environment
temperature is too
high or too low
Please
re-measure
following the
product instruction
9. Product technical parameters
05 06 07
Not to be disposed of as ordinary waste
The
screen
display
" "
10. Warranty card
Warranty card
Product name:
Infrared thermometer
Model:
YT-1
●Device main structure and composition
The thermometer consists of a housing, a sensor, a
display, and a circuit board.
4. Matters need attention
3. The symbols related to safety requirements
in this device and their meanings:
●Measure temperature of human body
1.Point the infrared thermometer probe at the center
of eyebrow, within 0-5cm of the forehead (do not
touch the forehead directly).
2.Press the Start measuring button ”. After
about 1 second, the infrared thermometer will have
vibration and sound prompts to display the
measurement results.
Notes:
①When the measurement
distance is more than
about 5 cm, the word "0-5"
will be displayed on the
screen, please close to the
subject, as the picture
shown at right;
②When the measurement distance is less than
about 5 cm, the device will automatically measure,
after the measurement, it will vibrate and have "di"
sound, and measurement result will be displayed.
③When temperature is 37.6 ℃ or higher, the device
will make quick continuous three-time "di-di-di"alarm
sound.
④when the measured result is beyond device
measurement range (32 ℃~43 ℃), thermometer will
make quick continuous three-time "di-di-di" alarm
sound, please make sure the measurement method
and the external environment is normal at this time.
[close to the
measurement indication]
●Sound on/off setting
When the prompt on/off button slides to " ", the
sound is off;
When the prompt on/off button slides to " ", the
sound is on.
●Power off
Long press Starting measurement button" " for
3~5 seconds to turn off the device;
②In the absence of any operation, the device will
automatically shut down after about 30 seconds.
Possible cause
Troubleshooting
methods
temperature
probe
ranging
probe
power on
measurement
button
prompt tone
key
battery
cover
LCD display
area
battery
reminder
distance unit
temperature unit
mute prompt
distance prompt
result display
10. If there is temperature difference compared
environment of the device storage with that of the
measuring environment, place the device in the
measuring environment for more than 30 minutes,
otherwise there may be error for the measurement
result.
11. When lens of sensor becomes dirty, please wipe
it gently with a soft dry cloth or cotton swab. Do not
wipe it with other material, otherwise it may cause
lens surface of senor abrasion or machine
malfunction.
12. Do not try to measure when the device is wet,
which may cause measurement results inaccurate.
13. Before measurement, please make sure that
there is no sweat, cosmetics or oil stains on the
forehead of subject.
14. Before measuring, please make sure that the
subject does not take a bath, exercise or have a meal
within 30 minutes, and the body is measured at a
steady state.
15. Before use, please keep sensor and cavities
clean.
16. During measurement, do not let the subject
directly face the air outlet of the sun, air conditioner or
heater, which will change the temperature of the
forehead. Please conduct measurement in a stable
environment as far as possible.
17. Under the condition which meets the requirement
of storage, the usage period of the device is 5
years(fragile and consumable parts are not included).
●Setting measure unit
In the shutdown state, long press the staring
measuring button “ ”for 10 seconds to enter the
temperature unit conversion state, and then short
press the staring measuring butto“ ” to select
”and“ ”temperature unit.
battery compartment and cover the battery cover.
Refer to the picuture at right:
cotton swab.
Disinfection of the product and its head: Gently wipe
the surface of the product and the head used for
measurement with a cotton swab dipped in a little
medical alcohol (70%), it can be used until the alcohol
has completely evaporated.
●Precaution for storage
This product should be stored in a dust-free, dry
place, please avoid direct sunlight; do not store in
places with high temperature, humidity, dust and
corrosive gas.
This product is a high-precision device, please do not
drop the device! Avoid drastic collisionsand jolts and
other adverse possibilities for transport.
If the probe or the product itself has been damaged,
do not continue to use it. Please do not use this
product for any purposes other than its intended use.
When it is used for children, please observe the
general safety precautions.
Environment temperature: 16℃~35
Relative humidity: ≤85%(RH)
Atmospheric pressure: 70kPa~106kPa
●Transportation and storage environment:
Environment temperature: -20℃~+55
Relative humidity: ≤85%(RH)
Atmospheric pressure: 70kPa~106kPa
Download manual in English (PDF, 1.28 MB)
(Consider the environment and only print this manual if it is really necessary)

Loading…

Rating

Let us know what you think about the Yuwell YT-1 Thermometer by leaving a product rating. Want to share your experiences with this product or ask a question? Please leave a comment at the bottom of the page.
Are you satisfied with the Yuwell YT-1 Thermometer?
Yes No
Be the first to rate this product
0 votes

Join the conversation about this product

Here you can share what you think about the Yuwell YT-1 Thermometer. If you have a question, first carefully read the manual. Requesting a manual can be done by using our contact form.

More about this manual

We understand that it’s nice to have a paper manual for your Yuwell YT-1 Thermometer. You can always download the manual from our website and print it yourself. If you would like to have an original manual, we recommend you contact Yuwell. They might be able to provide an original manual. Are you looking for the manual of your Yuwell YT-1 Thermometer in a different language? Choose your preferred language on our homepage and search for the model number to see if we have it available.

Specifications

Brand Yuwell
Model YT-1
Category Thermometers
File type PDF
File size 1.28 MB

All manuals for Yuwell Thermometers
More manuals of Thermometers

Frequently Asked Questions about Yuwell YT-1 Thermometer

Our support team searches for useful product information and answers to frequently asked questions. If you find an inaccuracy in our frequently asked questions, please let us know by using our contact form.

I just took my temperature but what is actually considered a 'healthy' body temperature? Verified

This partially depends on the body part where the temperature was taken. The following can be considered healthy: Forehead 35,8 to 37,6 °C, Ear 36 to 37,8°C, rectally 36,3 to 37,8°C and orally 36 to 37,4°C.

This was helpful (952) Read more

What parts of the body are best to measure body temperature? Verified

This depends on the thermometer. Some thermometers are designed for use on specific parts of the body. With a regular thermometer, rectal usage is the fastest and most accurate way to measure body temperature.

This was helpful (330) Read more

From what temperature does a person officially have a fever? Verified

A temperature above 38°C is officially considered a fever.

This was helpful (245) Read more
Manual Yuwell YT-1 Thermometer

Related products

Related categories