Cleaning
Maintenance
Troubleshooting
Specification
1. Please clean the surface of the device before using.
Wipe the device with medical alcohol (70% isopro-
pyl alcohol)first, and then let it dry in air or clean it
by dry clean fabric. When cleaning the device with
water, the water temperature should be lower than
60°C
2. Using the medical alcohol to disinfect the product
after use, prevent from cross infection for next time
use.
3. The best storage environment of the device is - 25°C
to 70°C ambient temperature and not higher than
90% relative humidity.
Note: 1. Do not sterilize, autoclave or immerse this de-
vice in liquid. Do not pour or spray any liquids
onto the device.
2. Do not use caustic or abrasive cleaning
agents, or any cleaning agent containing am-
monium chloride or isopropyl alcohol.
Recommends user to return this device to the man-
ufacturer perform the following checks every 24
months.
- Inspect the equipment for mechanical and function-
al damage or deterioration.
- Ensure all user interface keys and accessories func-
tion normally.
Note: Manufacturer use Index2 SpO2 simulator to
verify operation of the pulse oximeter device.
Symptoms Check points Corrections
SpO2 or
pulse rate
cannot
displayed
Applied nger
improperly.
Place the nger prop-
erly and try again.
SpO2 is too low to
detect
Try again; go to con-
sult with your physi-
cian if you are sure the
device works well.
SpO2 or
pulse rate are
not displayed
stably
Applied nger
improperly.
Place the nger prop-
erly and try again.
Finger is shaking
or body is moving.
keep body steady
No display
when button
is pressed
Batteries run
down
Replace with new
batteries
Batteries not
inserted correctly.
Re-insert batteries
The display
disappears
suddenly
The device will
auto power o
when it gets no
signal.
Normal
Low battery
Replace with new
batteries
Note: If the unit does not work, return it to your
dealer. Under no circumstance should you
disassemble and repair the unit by yourself.
SpO2
Measuring
range
35%~99%, (the resolution is 1%).
Accuracy 70%~99%: ±2%, Below 35~69%:
unspecied.
Optical Sensor The wavelength of red LED is 660 nm
and Infrared LED is 905/880 nm with
maximum optical output power of 4
mW/sr.
Pulse
Measuring
range
30 bpm~250 bpm (the resolution is 1
bpm)
Accuracy ±3 bpm
Power source AAA × 2 (Alkaline)
Battery life Continually for 16 hours with two alkaline
batteries
Operating
Condition
Temperature: 5°C~35°C (41°F ~ 95°F),
Relative Humidity: 15-95% (non
condensing), Atmospheric pressure:
700hPa ~ 1060hPa, Attitude: -1,280 to
12,000 feet (-390m to 3,658m)
Storage /
Transportation
Condition
Temperature: -25°C~+70°C(-13°F ~
158°F), Relative humidity: 15-90%(non
condensing), Atmospheric pressure:
700hPa ~ 1060hPa, Attitude: -1,280 to
12,000 feet (-390m to 3,658m), The time
from 70°C or -25°C back to use : 3 hours
Dimensions 61.3(L) × 31W) × 36.7(H) mm
Weight About 33g (without the batteries)
Standards IEC60601-1-2, Class B, IEC60601-1, Type BF,
ISO80601-2-61, IEC/EN60601-1-11
Type BF applied parts
IP Classication IP22: Protection against harmful ingress of
water and particulate matter
Note:
- A description of the effect on displayed and transmit-
ted SpO2 and pulse rate:
- Data averaging: 4 seconds for SpO2; 8 seconds for
pulse rate.
- Data update delay: Less than 2 seconds.
be evaluated at the measurement site. Rather, they
are exclusively used to display the current signal vari-
ation at the measurement site and do not enable re-
liable diagnostics for the pulse.
- A warning that other cables and accessories may
negatively aect EMC performance.
- A warning regarding stacking and location close to
other equipment.
- A warning that use of other accessories results in
non-compliance.
- The safety way in high-temperature environment for
all people use is measuring for 10 minutes, and turn
it o for 20 minutes before measure again.
- The oximeter is calibrated in the factory before sale,
there is no need to calibrate it during its life cycle.
Model: SD100
www.rossmax.com
Fingertip Pulse Oximeter
Rossmax InnoTek Corp.
12F., No. 189, Kang Chien Rd., Taipei,
114, Taiwan.
CMC Medical Devices & Drugs S.L.
C/ Horacio Lengo Nº 18, CP 29006,
Málaga, Spain
WARNING: The symbol on this product means that it's an electronic product
and following the European directive 2012/19/EU the electronic products have
to be dispose on your local recycling centre for safe treatment.
EMC guidance and manufacturer’s declaration
Recommended separation distances between portable and mobile RF communications equipment and the ME equipment
The Finger-tip pulse oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the Finger-tip pulse oximeter can help prevent electromagnetic interference by maintaining a
minimum distance between portable and mobile RF communications equipment (transmitters) and the Finger-tip pulse oximeter
as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz , d=[3.5/
V1]√P
80 MHz to 800 MHz , d=[3.5/
E1]√P
800 MHz to 2,5 GHz , d=[7/
E1]√P
0.01 0.1 0.1 0.2
0.1 0.4 0.4 0.7
1 1.2 1.2 2.3
10 3.7 3.7 7.4
100 11.7 11.7 23.3
Declaration – electromagnetic emissions
The Finger-tip pulse oximeter is intended for use in the electromagnetic environment specied below. The customer or the user of
the Finger-tip pulse oximeter should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 Portable and mobile RF communications equipment should be used no
closer to any part of the EQUIPMENT or SYSTEM including cables, than the
recommended separation distance calculated from the equation applicable
to the frequency of the transmitter. Interference may occur in the vicinity of
equipment marked with the following symbol.
RF emissions CISPR 11 Class B The Finger-tip pulse oximeter is suitable for use in all establishments,
including domestic establishments and those directly connected to the
public low-voltage power supply network that supplies buildings used
for domestic purposes.
Harmonic emissions IEC 61000-3-2 N/A
Voltage uctuations/Flicker emissions
IEC 61000-3-3
N/A
Declaration – electromagnetic emissions and immunity – for EQUIPMENT and SYSTEMS that are use in the professional health-
care facility environment or in the home healthcare environment
The Finger-tip pulse oximeter declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Conducted RF IEC
61000-4-6
3 Vrms ; 6 Vrms ; 150
kHz to 80 MHz
N/A Portable and mobile RF communications equipment
should be used no closer to any part of the EQUIPMENT
or SYSTEM including cables, than the recommended
separation distance calculated from the equation ap-
plicable to the frequency of the transmitter. Interference
may occur in the vicinity of equipment marked with the
following symbol.
Radiated RF IEC
61000-4-3
3 V/m ; 10V/m ; 80
MHz – 2.7 GHz l 80%
3 V/m ; 10V/m ; 80
MHz – 2.7 GHz ; 80%
Proximity elds
from RF wireless
Communications
equipment IEC
61000-4-3
27 V/m 385 MHz 27 V/m 385 MHz
28 V/m 450 MHz 28 V/m 450 MHz
9 V/m 710 MHz 9 V/m 710 MHz
745 MHz 745 MHz
780 MHz 780 MHz
28 V/m 810 MHz 28 V/m 810 MHz
870 MHz 870 MHz
930 MHz 930 MHz
28 V/m 1720 MHz 28 V/m 1720 MHz
1845 MHz 1845 MHz
1970 MHz 1970 MHz
28 V/m 2450 MHz 28 V/m 2450 MHz
9 V/m 5240 MHz 9 V/m 5240 MHz
5500 MHz 5500 MHz
5785 MHz 5785 MHz
Declaration – electromagnetic immunity
The Finger-tip pulse oximeter system is intended for use in the electromagnetic environment specied below. The customer or
the user of the Finger-tip pulse oximeter system should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level Electromagnetic environment - guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV,
±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8
kV, ±15 kV air
Floors should be wood, concrete or ceramic tile.
If oors are covered with synthetic material, the
relative humidity should be at least 30 %.
Electrical fast transient/
burst IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
N/A Mains power quality should be that of a typical
commercial or hospital environment.
Surge IEC 61000-4-5 ±0.5 kV
±1 kV dierential mode
±2 kV common mode
N/A Mains power quality should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines IEC
61000-4-11
0 % U
T
; 0, 5 cycle At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°
0 % U
T
; 1 cycle And 70
% U
T
; 25/30 cycle Single
phase: at 0°
N/A Mains power quality should be that of a typical
commercial or hospital environment. If the user
of the EQUIPMENT or SYSTEM requires continued
operation during power mains interruptions, it is
recommended that the EQUIPMENT or SYSTEM
be powered from an uninterruptible power sup-
ply or a battery.
Power frequency (50/60
Hz) magnetic eld IEC
61000-4-8
30 A/m 30 A/m Power frequency magnetic elds should be at
levels characteristic of a typical location in a typi-
cal commercial or hospital environment.
Warranty Card
This instrument is covered by a 2 years guarantee from the date of purchase,
batteries and accessories are not included. The guarantee is valid only on
presentation of the guarantee card completed by the dealer confirming date
of purchase or the receipt. Opening or altering the instrument invalidates
the guarantee. The guarantee does not cover damage, accidents or non-
compliance with the instruction manual. Please contact your local seller/dealer
or www.rossmax.com.
Customer Name: _______________________________
Address: ____________________________________
Telephone: __________________________________
E-mail address: ________________________________
Product Information: ___________________________
Date of purchase: ______________________________
Store where purchased:
___________________________________________
IN0SD100000000010
RI_IB_SD100_EN_
TP_ver2012
RI_IB_SD100_EN_TP_ver2012.indd 1 2020/9/25 下午 04:19:14
Download manual in English (PDF, 1.41 MB)
(Consider the environment and only print this manual if it is really necessary)

Loading…

Rating

Let us know what you think about the Rossmax SD100 Pulse Oximeter by leaving a product rating. Want to share your experiences with this product or ask a question? Please leave a comment at the bottom of the page.
Are you satisfied with the Rossmax SD100 Pulse Oximeter?
Yes No
Be the first to rate this product
0 votes

Join the conversation about this product

Here you can share what you think about the Rossmax SD100 Pulse Oximeter. If you have a question, first carefully read the manual. Requesting a manual can be done by using our contact form.

More about this manual

We understand that it’s nice to have a paper manual for your Rossmax SD100 Pulse Oximeter. You can always download the manual from our website and print it yourself. If you would like to have an original manual, we recommend you contact Rossmax. They might be able to provide an original manual. Are you looking for the manual of your Rossmax SD100 Pulse Oximeter in a different language? Choose your preferred language on our homepage and search for the model number to see if we have it available.

Specifications

Brand Rossmax
Model SD100
Category Pulse Oximeters
File type PDF
File size 1.41 MB

All manuals for Rossmax Pulse Oximeters
More manuals of Pulse Oximeters

Manual Rossmax SD100 Pulse Oximeter

Related products

Related categories